Res. Biomed. Eng. (Online) ; 33(1): 42-49, Mar. 2017. tab, graf
Article in English | LILACS | ID: biblio-842478

ABSTRACT

Abstract Introduction Therapeutic ultrasound (TUS) is a widespread modality in physiotherapy, and the water bag technique is a coupling method employed in the presence of anatomical irregularities in the treatment area. The aim of the present study is to evaluate the acoustic attenuation of the water bag and its effectiveness as a TUS coupling agent. Methods The rated output powers (ROPs) of the TUS equipment were evaluated based on IEC 61689. Then, a radiation force balance was used to measure ROP with and without a water bag (latex and nitrile gloves filled with deionized water) between a TUS transducer and the cone-shaped target of the balance. Each experiment was performed five times for each nominal power (0.5, 1.0, 2.0, 3.0, 4.0, 5.0, 6.0, and 7.0 W) and in the following configurations: without the water bag (A), with nitrile gloves and with (B) and without (C) a height controller, and latex gloves with (D) and without (E) height controller. ROPs obtained in different media were compared. Results The highest relative error of ROP was 16.72% for 0.5 W. Although the power values of the equipment were within the range recommended by IEC, there was a significant difference between the ROP values measured with A and with B, C and D. Conclusion As intensity differences below 0.5 W/cm2 are considered clinically not relevant, conditions A, B, C, D, or E can be used interchangeably.